FDA Recall: Landauer — Class II
20240103
MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction | Product: Landauer microSTARii reader, Medical D...
View Source
FDA Device Recall: microStar Reader, Model Numbers: 10015-000 (marketed in EMEA as 100, 102, 110,
2024-03-27
Class II — Permanent discontinuance of the microSTARii medical dosimetry system. Landauer has initiated this action to remove all reference to use of the microStar readers with the nanoDot or any oth...
View Source
FDA Device Recall: Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000
2024-01-03
Class II — MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with t...
View Source
FDA Recall: Landauer — Class II
20230920
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDo | Product: nanoDot D2DNS, Model Numbers: a) 18...
View Source
FDA Recall: Landauer — Class II
20230920
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDo | Product: nanoDot D2DNS, Model Numbers: a) 03...
View Source
FDA Recall: Landauer — Class II
20230920
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDo | Product: Constancy (80 kVp) nanoDot D2DNN, Mode...
View Source
FDA Recall: Landauer — Class II
20230920
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDo | Product: QC (80 kVp) screened nanoDot D2DNS, Mo...
View Source
FDA Recall: Landauer — Class II
20230920
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDo | Product: nanoDot A - item 1 of 3 for 03500-000,...
View Source
FDA Recall: Landauer — Class II
20230920
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDo | Product: Reader, nanoDot adapter, EU Model Numb...
View Source
FDA Recall: Landauer — Class II
20230920
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDo | Product: Calibrate (80 kVp) nanoDot D2DNS, Mode...
View Source
FDA Recall: Landauer — Class II
20230920
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDo | Product: nanoDot D2DNS, for custom calibrate, O...
View Source
FDA Recall: Landauer — Class II
20230920
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDo | Product: nanoDot kit (contains nanoDot A, B, C)...
View Source
FDA Recall: Landauer — Class II
20230920
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDo | Product: nanoDot D2DNS, for custom calibrate, O...
View Source
FDA Device Recall: nanoDot D2DNS, Model Numbers: a) 18100-000; b) 18100-1MO; c) 18100-2MO;
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: nanoDot D2DNS, Model Numbers: a) 03057-1MO; b) 03057-3MO; c) 03057-OTO;
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: Constancy (80 kVp) nanoDot D2DNN, Model Numbers: a) 18150-000 (EMEA number VK
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: QC (80 kVp) screened nanoDot D2DNS, Model Number 03066-OTO; radiation monitoring
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: nanoDot A - item 1 of 3 for 03500-000, Model Number 03501-000; radiation monitor
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: Reader, nanoDot adapter, EU Model Number VINLADA003; radiation monitoring dosime
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: Calibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 18120-000; b) 18120-KIT
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: nanoDot D2DNS, for custom calibrate, OTO, Model Numbers 03034-OTO; radiation mon
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: nanoDot kit (contains nanoDot A, B, C), Model Number 03500-000; radiation monito
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: nanoDot D2DNS, for custom calibrate, OTO, Model Number 03033-OTO; radiation moni
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: QC (Cs-137) nanoDot D2DNS, Model Numbers: a) 04225-000 (EMEA number VINLNAN00
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: nanoDot D2DNN, Model Numbers: a) 03053-1MO; b) 03053-1WK ; c) 03053-3MO;
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: nanoDot D2DNN, Model Numbers: a) 18105-000; b) 18105-KIT; radiation monitor
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: QC (80 kVp) nanoDot D2DNS, Model Numbers: a) 18130-000; b) 18130-KIT; c)
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: nanoDot D2DNS, Model Numbers: a) 03056-000 (EMEA number VINLNAN002); b) 03
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: QC (Cs-137) screened nanoDot D2DNS, Model Number 03067-OTO; radiation monitoring
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: NanoDots, Model Number 04297-000; radiation monitoring dosimeter used with the m
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: Constancy (Cs-137) nanoDot D2DNN, Model Numbers: a) 18155-000 (EMEA number VK
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: Calibrate (Cs-137) screened nanoDot D2DNS, Model Number 03065-OTO; radiation mon
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: nanoDot D2DXS, Model Numbers: a) 03063-1MO; b) 03063-3MO; c) 03063-OTO;
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: Calibrate (80 kVp) screened nanoDot D2DNS, Model Number 03064-OTO; radiation mon
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: QC (Cs-137) nanoDot D2DNS, Model Numbers: a) 18135-000; b) 18135-KIT; c)
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: nanoDot D2DXN, Model Numbers: a) 03061-1MO; b) 03061-3MO; c) 03061-OTO;
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: nanoDot B - item 2 of 3 for 03500-000, Model Number 03502-000; radiation monitor
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: Calibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 18121-000; b)18121-KIT;
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
FDA Device Recall: nanoDot D2DXS, Model Numbers: a) 03062-1MO; b) 03062-3MO; c) 03062-OTO;
2023-09-20
Class II — LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potentia...
View Source
ny_corps
2 records
NY Corp: LANDAUER PARTNERS LLC
2014-02-05
DOMESTIC LIMITED LIABILITY COMPANY | County: New York | Jurisdiction: New York
View Source
NY Corp: LANDAUER, INC.
2011-11-21
FOREIGN BUSINESS CORPORATION | County: New York | Jurisdiction: Delaware
View Source
Be informed when major changes happen on C.F.A.I. — new data sources, platform updates, critical enforcement alerts
No spam. Only major updates. Unsubscribe anytime.